For research as well as for external services, the DM & QC research group at PMA relies on two metrology labs, located in adjacent rooms. The interferometry room has a volume of approximately 105 m³. The large laboratory is approximately 4 times larger. Each room has its own air conditioning unit, designed to keep the temperature in the rooms at a constant 20°C. Disturbances are, among others, persons in the room, illumination, machines. The foundations supporting the metrology rooms are distinct from the rest of the building, where offices and other labs at PMA are located. The negative influence of vibrations caused by machines within and outside the other sections of the building are hereby reduced. Neither room has windows, this to avoid influence of direct sunlight. The metrology labs are accessed via an air-lock. This avoids air draughts while persons enter or leave these facilities.
![]()
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|